Calipered Kinematic Alignment Verification Form

Brought to you by Ortho Research Master® and David F. Scott, MD, of Spokane Joint Replacement Center.

Ortho Research Master Spokane Joint Replacement Center
Saved on this deviceSaved

Case

Operative side
Fixed flexion contracture
Additional distal resection
Wear pattern

Distal femoral resection

Target thickness is for the unworn condition, after accounting for the 1 mm saw blade kerf and any fixed flexion contracture. Worn condyles subtract 2 mm.

Target (unworn)

Posterior femoral resection

Target thickness is for the unworn condition, after accounting for the 1 mm saw blade kerf. A worn condyle uses a 2 mm shim, which is added to the measured thickness, so the target stays the same.

Target (unworn)

Tibial resection

Target: equal thickness measured at the base of the tibial spines. Differences use the after-recut value when one is entered.

Anterior Δ
—
Posterior Δ
—
Average Δ
—
PCL condition
Tibial recuts
Flat recut amount

Patella resurfacing

Resurfaced
Pre-cut thickness
Post-cut thickness
Patella resection
—

Ligament release

Ligament release performed

Implant info

Insert type

Export this case

Contact us

Questions, feedback or feature requests are welcome. Please don’t include any patient information.

Start a new case?

This clears every field on this device. Export the current case first if you need it.

Data responsibility & HIPAA notice

This tool can capture information that may be considered Protected Health Information (PHI). By proceeding, you acknowledge that you are solely responsible for any PHI entered, stored, exported, or otherwise processed using this tool, and that you will handle it in compliance with applicable privacy and security laws, including HIPAA where applicable. The hosting website and its operators do not assume liability for your use or handling of any data.

This notice is informational only and does not constitute legal advice. Consult your compliance officer or legal counsel.